- First turn on M6
- Setup screen M6
- Protective eyewear M6
- Password overview M6
Controls and warning devices
- The START/STOP button (A);
- The red emergency button (B);
- Laser warning lamp (C);
- Operator display (D);
- The connector of the handpiece (E);
- General switch and power socket (F);
- Interlock connector and repeat of the laser emission visual signal plug (G);
To access the appliance, connect the power supply connector supplied to the appliance ‘power supply’ socket. Press the START/STOP button C releasing immediately you hear a beep.
Switching on the device
To turn on the device connect the connector of the supplied power supply unit to the “power supply” socket on the device.
Turn on the main switch.
Briefly press the START / STOP button until you hear a beep.
Switching off the device
To switch the appliance off, press and hold the START/STOP button until you hear a series of beeps.
Emission warning light
The blue emission warning light emits a flashing light signal when Laser emission is standing-by, a constant light when Laser emission is enabled and with the handpieces connected a Laser emission is possible.
Acoustic signal
The device is equipped with an audible alarm to signal the beginning, the execution of the treatment, the interruption or the end of the treatment.
When turn on
After the on switch is pressed, the software will load and a warning sign will remind the user about safety.
Always use the provided stylus to touch the screen.
Press the arrow to close the warning screen and access the HOME screen.
Password screen
- The user will be asked to input a password.
- Enter the password that you were provided, either by email or on a USB stick then press OK.
- The default password of the device , when none has been entered, is 1111.
- A – ERASE KEY
- B – CONFIRM KEY (OK)
- C – NUMERICAL ENTRY
Home screen
After entering the password, the “home” screen is reached.
A – Sections on the human body for access to ‘Pain management’ mode
B – Access to appliance set-up
C – Access to personal programs
D – Access to anti-oedema program
E – Access to biostimulation program
F – Direct setting of treatment parameters