- Unpack M6
- Assemble parts M6
- Care of unit M6
- Room setup M6
- Device controls M6
Click here for M6 User Guide
Controls and warning devices
- The START/STOP button (A);
- The red emergency button (B);
- Laser warning lamp (C);
- Operator display (D);
- The connector of the handpiece (E);
- General switch and power socket (F);
- Interlock connector and repeat of the laser emission visual signal plug (G);
To access the appliance, connect the power supply connector supplied to the appliance ‘power supply’ socket. Press the START/STOP button C releasing immediately you hear a beep.
Switching on the device A
To access the appliance, connect the power supply connector supplied to the appliance ‘power supply’ socket. Press the START/STOP button, releasing immediately you hear a beep.
Switching off the device A
To switch the appliance off, press and hold the START/STOP button until you hear a series of beeps.
Emergency Stop B
The emergency stop should only be used in emergency cases to interrupt immediately the laser beam emission.
The switch has only two positions:
UP (or OUT) – The device is supplied with power and can emit laser radiation;
DOWN (or IN) – The laser emission is interrupted
Need to pull the plunger out to restart machine
Emission warning light C
The blue emission warning light emits a flashing light signal when Laser emission is standing-by, a constant light when Laser emission is enabled and with the handpieces connected a Laser emission is possible.
Acoustic signal
The device is equipped with an audible alarm to signal the beginning, the execution of the treatment, the interruption or the end of the treatment.
Handpiece connection E
Connect the handpiece under the main console.
Always store the handpiece safely in the holder on the side of the unit.
Power cord F
The power supply is connected to the device by a cable to socket F at the base of the unit. It is connected to the wall socket by a standard 3-pin IEC cable as frequently used in computers. The provided cable may be swapped out for a longer cable or added to with a IEC extension cable.