Handpiece
- A – Treatment indication: Biostimulating
- B – Treatment area setting
- C – Indication of energy density
In Biostimulating mode the user can change or set the emission mode, frequency, time and intensity values as well as the size of the area to be treated.
The user can also optimise therapeutic application considering the patient’s phototype: fair phototype (A) and dark phototype (B).
Adjust the treatment area by pressing the “+” or “-” buttons or select half way between the two to open a numeric keypad. Enter the area required and press OK. The device will alter the settings based on the area entered.
The treatment envisages the user scanning the handpiece over the area being treated.
Scanning involves slowly moving the handpiece across the body part touching the skin.
When ready, the operator presses START.
After pressing START an intermittent warning buzzer sounds, a warning lamp flashes and a warning triangle appears on the display to warn that laser is armed and ready to emit.
By pressing the button on the handpiece it then begins supplying MLS laser radiation.
The supply continues until the time expires when the device will stop and revert to the previous screen.
Robotic
In Biostimulating mode the user can change or set the emission mode, frequency, time and intensity values as well as the size of the area to be treated.
Adjust the treatment area size by pressing the “+” or “-” buttons on the X and Y axes. The device will alter the settings based on the area entered.
The treatment envisages the user directing the head 20cm away from the patient. The head will need to be moved if large areas are treated.
When ready, the operator presses START.
After pressing START an intermittent warning buzzer sounds, a warning lamp flashes and a warning triangle appears on the display to warn that laser is armed and ready to emit.
At the end of the warning message, pressing the
button, the robotised head starts to emit laser radiation. In this state, the device shows a screen stating the treatment values, residual time and the cumulative value of energy emitted.
The supply continues until the time expires when the device will stop and revert to the previous screen.